A–4
VM100 & VM101 User Manual
Table A–3: Measurement accuracy (cont.)
Category
1
Description
Hum
2
REQ: ±1.0% for hum ≤4.0%
±2.0% for hum >4.0% and <15.0%
RI: Typically ±0.5% for hum ≤4.0%
Signal-to-Noise Ratio
(Weighted or Flat)
REQ: ±1 dB, 28 dB to <45 dB
±2 dB, 45 dB to <60 dB
±3 dB, 60 dB to 67 dB (WGT)
RI: VM100: Noise is weighted with EIA
RS-250-B 5 MHz weighted noise
filter. (Tektronix part number
015-0215-00 equivalent.)
VM101: Noise is weighted with
CCIR 567 weighted noise filter
Group Delay REQ: <20 ns
RI: 50 kHz to 4.8 MHz, within the
bandwidth of the signal under test
Sync Amplitude
2
REQ: <1.0%
Burst Amplitude REQ: <1.0%
Bar Amplitude REQ: <1.0%
Luminance Nonlinearity REQ: <1.0%
1
All tests are performed with AVG enabled. Processing time for each
measurement is a function of the signal-to-noise ratio.
2
The hum and sync amplitude measurements are inaccurate in the presence
of Sound In Sync (SIS) signals.