Automated Video Measurement Set — VM6000
Physical Characteri stic s
Benchtop Configuration
Dimension
mm
in.
Height 292 11.48
Width 451 17.75
Depth 265 10.44
Weight kg lb.
Net 15 32
Shipping
28.9 63.75
Rackmount Configuration
Dimension
mm
in.
Height 323 12.25
Width 479 18.85
Depth (from rackmounting
ear to back of instrument)
231.75 9.12
Weight kg lb.
Net 17.4 37.5
Kit 2.5
5.5
Mechanical
Cooling – Required Clearance
Dimension
mm
in.
Top 0 0
Bottom 0 0
Left side
00
Right side 76 3
Front 0 0
Rear 0 0
Environmental
Characteristic
Description
Temperature
Operating +10 °C to +45 °C
Nonoperating
-40°Cto+71°C
Humidity
Operating 5% to 95% relative humidity (RH) with a maximum
wet bulb temperature of +29 °C at or below +50 °C,
noncondensing. Upper limit derated to 45% RH above
+30°Cupto+50°C
Nonoperating
5% to 95% relative humidity (RH) with a maximum
wet bulb temperature of +29 °C at or below +60 °C,
noncondensing. Upper limit derated to 45% RH above
+30°Cupto+50°C
Altitude
Operating 10,000 ft. (3,048 m)
Nonoperating
40,000 ft. (12,190 m)
Random Vibration
0.000125 G
2
/Hz from 5 to 350 Hz
-3 dB/octave from 350 to 500 Hz
0.0000876 G
2
/Hz at 500 Hz
Operating
Overall level of 0.27 G
RMS
0.0175 G
2
/Hz from 5 to 100 Hz
-3 dB/octave from 100 to 200 Hz
0.00875 G
2
/Hz from 200 to 350 Hz
-3 dB/octave from 350 to 500 Hz
0.006132 G
2
/Hz at 500 Hz
Nonoperating
Overall level of 2.28 G
RMS
Regulatory
Electromagnetic
Compatibility
93/68/EEC; EN61326:1997 +A1 1998+A2:2000
Certifications UL 3111-1, CSA1010.1, ISO11469,EN61010-1, IEC
61010-1